40ng MOP Drug Abuse Test Kit For Opiates Detection In whole blood/serum/plasma

Basic Information
Place of Origin: China
Brand Name: AllTest
Certification: CE, ISO13485
Model Number: Cassette (DMO-402)
Minimum Order Quantity: N/A
Price: Negotiation
Packaging Details: 40 cassette/box
Delivery Time: 2-4 Weeks
Payment Terms: Ex works
Supply Ability: 10 M tests/month
Format: Cassette Specimen: WB/S/P
Kit Size: 40T/Kit Storage: 2-30℃
Color: White Certificate: CE
High Light:

one step rapid test

,

home drug test kits

40ng MOP Drug Abuse Test Kit For Opiates Detection In whole blood/serum/plasma

Applications:

 

The MOP Rapid Test Cassette (whole blood/serum/plasma) is a lateral flow chromatographic immunoassay for the detection of Morphine in whole blood or serum or plasma at a cut-off concentration of 40ng/ml. This test will detect other related compounds, please refer to the analytical specificity table in this package insert.
This assay provides only a qualitative, preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

 

Description:  

 

Opioid analgesics comprise a large group of substances which control pain by depressing the CNS. Large doses of morphine can produce higher tolerance levels, physiological dependency in users, and may lead to substance abuse. Morphine is excreted unmetabolized, and is also the major metabolic product of codeine and heroin. Morphine is detectable in the whole blood or serum or plasma for several days after an opiate dose1. The MOP Rapid Test Cassette is a rapid whole blood/serum/plasma screening test that can be performed without the use of an instrument. The test utilizes a monoclonal antibody to selectively detect elevated levels of Morphine in whole blood/serum/plasma. The MOP Rapid Test Cassette yields a positive result when Morphine in whole blood/serum/plasma reaches 40ng/ml.

 

How to use? 

 

Allow the test, specimen, buffer and/or controls to reach room temperature (15-30°C) prior to testing.
For serum or plasma specimen:
1. Bring the pouch to room temperature (15-30℃) before opening it. Remove the cassette from the sealed pouch and use it within one hour.
2. Place the cassette on a clean and level surface. Hold the dropper vertically and transfer 1 full drop of serum or plasma (approximately 40μl), then add 2 drops of buffer (approximately 80μl) to the specimen well of the cassette, and then start the timer. Avoid trapping air bubbles in the specimen well. See illustration below.
3. Wait for the colored line(s) to appear. Read the result at 5 minutes. Do not interpret the result after 10 minutes.
 
40ng MOP Drug Abuse Test Kit For Opiates Detection In whole blood/serum/plasma 0
 
1 Bring the pouch to room temperature before opening it. Remove the test cassette from the sealed pouch and use it as soon as possible.
2 Place the cassette on a clean and level surface.
For Venipuncture Whole blood specimen:
Hold the dropper vertically and transfer 1 drop of whole blood (approximately 40μl) to the specimen well, then add 2 drops of buffer (approximately 80μl), and start the timer. See illustration below.
For Fingerstick Whole blood specimen:
To use a capillary tube: Fill the capillary tube and transfer approximately 40μl of fingerstick Whole blood specimen to the specimen area of test cassette, then add 2 drops of buffer(approximately 80μl) and start the timer. See illustration below.
3 Wait for the colored line(s) to appear. Read results at 5 minutes. Do not interpret the result after 10 minutes.
 
 
40ng MOP Drug Abuse Test Kit For Opiates Detection In whole blood/serum/plasma 1
 

 

 

INTERPRETATION OF RESULTS


(Please refer to the illustration above)


NEGATIVE:* Two lines appear. One colored line should be in the control line region (C), and another apparent colored line should be in the test line region (T). This negative result indicates that the Morphine concentration is below the detectable cut-off level.


*NOTE: The shade of color in the test line region (T) may vary, but it should be considered negative whenever there is even a faint colored line.
POSITIVE: One colored line appears in the control line region (C). No line appears in the test line region (T). This positive result indicates that the Morphine concentration exceeds the detectable cut-off level.


INVALID: Control line fails to appear. Insufficient specimen volume or incorrect procedural techniques are the most likely reasons for control line failure. Review the procedure and repeat the test with a new test. If the problem persists, discontinue using the test kit immediately and contact your local distributor.