The AllCheckTM blood glucose monitoring system will give you accurate readings from fresh capillary whole blood. This product is for professional use in medecal institutes or for self-testing by diabetics. Test results serve only as helpful information, not as proft for clinical diagnoses.
Specification |
|
|
Cat. No. |
Product Description |
|
Product Namme |
AllChekTM Blood Glucose Meter |
|
Size |
83.6mm (L) x 51.6mm (W) x 15.3mm (Thickness) |
|
Glucose Test Range |
1.1 - 33.3 mmol/L (20 - 600mg/dL) |
|
Results Dispaly |
Plasma quivalent |
|
Minimum sample volume |
0.6μL |
|
Test time |
5 seconds |
|
Battery |
1 x CR 2032 3.0V coin cell battery |
|
Battery lige |
>1,000 radings |
|
Glucose Concentration Units |
mmol/L depending on the standard of your country. |
|
Memory storage |
500 test results with date and time stamp |
|
Auto shutdown |
Automatci shutdown after 2 minites |
|
Display size |
40mm x 42mm |
|
Weight |
About 50 grams (including battery) |
|
Operating temperature |
5 - 45℃ (41-113℉) |
|
Operating humidity |
20 - 90% (non-condensing) |
|
Hematocrit range |
30 - 55% |
Product Components
Depending on which AllCheckTM product you purchased, some of the components may to be purchased seperately. Please check the list of components printed on the box for details on which components are included with your purchase.
Blood Glucose Meter: The meter reads the test strips and displays the blood glucoseconcentration. Use this diagram to become familiar with all of theparts of your meter.
Test strips: AlIChekTM blood glucose test strips contain chemical reagents. Afterthe test strip is placed into the test strip port and a blood sample isapplied, the blood automatically wicks into test window. A transientelectrical current is generated, and this current is measured todetermine the correct blood glucose level reading.
Control solution: Control solution is a glucose solution of known concentration that is usedto confirm that your ÃlChek"' glucose meter and test strips are workingproperly. It is important to run a quality controll test regularly to make surethat you are geting accurate results.
Testing blood glucose:
1. Insert the test strip all of the way into the test strip port, and the meter will turn on automatically. If the audio is on, you will hear a "beep"sound, and all of the icons on the display will turn on at the same time.If the strip is inserted upside down, the meter will not turn on.
2. The test strip has been inserted correctly when you see the test strip icon and flashing blood drop icon. If the test strip is not correctly inserted, the meter will not turn on - repeat the steps above.
3. Apply blood to the sample area of the test strip. If the audio is on, the meter will beep, telling you that enough blood has been applied and the meter will start the measurement.
DO NOT:·
l Apply blood to the front or back of the test strip.·
l Smear the blood drop onto the test strip.
l Press your finger against the test strip.
4. The display will count down 5 seconds during the measurement process. If the audio is on, the test will end with a beep. If you applied a bloodsample but the meter does not begin a countdown, you may reapply a second drop of blood within 3 seconds.
If you press the ">" button, the test results will be marked with the "#" icon. Results marked with "#" symbol will not be used in the 7, 14, 30,60 or 90 day average calculations. If you accidentally mark your result with the "#" icon, simply press the ">" button again to cancel the mark. After marking a result with "#", please test again with a new strip.
If an error code is displayed, please check the troubleshotinginstructions. If "HI' or "LO" symbol is displayed, refer to the"HI" and "LO" Messages section.
5. Record valid results in your logbook with date and time, and compare them to the target goals set by your healthcare professional. (Refer to Suggested Testing Times and Target Goals in Instruction Manual)
6. After the test has been completed, slide the test strip ejector to pop out the test strip, and the glucose meter will automatically turn off.
Ordering Information |
|
Cat. No. |
Product Description |
|
AGL-ST10/ST25-a |
AllChekTM Glucose Meter Set, with Start Kit 10pcs/25pcs, Results formula: mmol/L |
|
AGL-ST20/ST25-b |
AllChekTM Glucose Meter Set, with Start Kit 10pcs/25pcs, Results formula: mg/dL |
|
AGL-NST-a |
AllChekTM Glucose Meter Set, without Start Kit, Results formula: mmol/L |
|
AGL-NST-b |
AllChekTM Glucose Meter Set, without Start Kit, Results formula: mg/dL |
|
AGL-S50 |
AllChekTM Glucose Strips, 25 strips into a canister, 50 strips into a box |
|
AGL-S100C |
AllChekTM Glucose Strips, 50 strips into a box |
|
AGL-IS50 |
AllChekTM Glucose Strip, 100 strips into a canister |
|
AGL-OST10/OST25 |
AllChekTM Glucose Start Kit 10 pcs/25pcs |
|
AGL-L50 |
Lancet, 50 pcs/box |
|
AGL-CS1/CS2/CSS |
AllChekTM Glucose Control Solution 1/2 set ( control solution 1&2) |