Tenamfetamine Drug Abuse Test Kit 500ng / Ml One Test Panel Powder Diagnosis

Basic Information
Place of Origin: China
Brand Name: AllTest
Certification: CE, ISO13485
Model Number: DMDA-X14
Minimum Order Quantity: N/A
Price: Negotiation
Packaging Details: 25 cassette/box
Delivery Time: 2-4 Weeks
Payment Terms: Ex works
Supply Ability: 10 M tests/month
Format: AllTest Specimen: Powder
Kit Size: 25 T/Kit Storage: 2-30℃
Cut-off: 500ng/ml Certificate: CE
High Light:

one step rapid test

,

home drug test kits

 

Tenamfetamine 500ng/ml One Test Panel Powder Diagnosis Kits Drug Abuse Test Kit With CE

 

Application:

 

The (±) 3,4-Methylenedioxy-Amphetamine Rapid Test Panel is a rapid chromatographic immunoassay for the qualitative detection of 3,4-Methylenedioxyamphetamine at a cut-off 500ng/ml.
This assay provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are indicated.

 

Description:  

 

3,4-Methylenedioxyamphetamine (MDA), also known as tenamfetamine (INN), or with the street name "Sally" or "Sass" or "Sass-a-frass", is a psychedelic and entactogenic drug of the
phenethylamine and amphetamine chemical classes. It is mainly used as a recreational drug, an entheogen, and a tool in use to supplement various types of practices for transcendence, including in meditation, psychonautics, and as an agent in psychedelic psychotherapy. It was first synthesized by G. Mannish and W. Jacobson in 1910. There are about 20 different synthetic routes described in the literature for its preparation.
The(±) 3,4-MethylenedioxyAmphetamine contained within the MDA Rapid Test Panel yields a positive result when the MDA concentration exceeds 500ng/ml.

 

How to use? 

 

Test device (in closed pouches), samples, and controls should be brought to room temperature (15-30°C) prior to testing. Do not open pouches until ready to perform the assay.
Remove the test device from its protective pouch and label the device with patient’s identification or control label.
FOR SURFACES
1.Wipe with the panel over the surface in which the drugs are expected.
2. Take off the cap of supplied tube;
3. Fill all buffers from the supplied tube of buffer into the protection cover.
4. Insert the Multi Test slowly and carefully into the protection cover with buffer
5. Wait for lines to appear on the membrane and read the results after 5 minutes and do not interpret the result after 10 minutes.

Tenamfetamine Drug Abuse Test Kit 500ng / Ml One Test Panel Powder Diagnosis 0


FOR SOLIDS
1. Open the tube and put the solid in to the buffer.
2. Close the tube with dropper and cap. Shake it a short time. Wait for 30 sec.
3. Take off the cap of supplied tube;
4. Fill all buffers with dissolved substances into the protection cover.
5. Insert the Multi Test slowly and carefully into the protection cover with buffer.
6. Wait for lines to appear on the membrane and read the results after 5 minutes and do not interpret the result after 10 minutes.

Tenamfetamine Drug Abuse Test Kit 500ng / Ml One Test Panel Powder Diagnosis 1

 

INTERPRETATION OF RESULTS
(Please refer to the illustration above)
Tenamfetamine Drug Abuse Test Kit 500ng / Ml One Test Panel Powder Diagnosis 2

NEGATIVE:* A colored line appears in the Control region (C) and colored lines appear in the Test region (T). This negative result means that the concentrations in the sample are below the designated cut-off levels for a particular drug tested.


*NOTE: The shade of the colored lines(s) in the Test region (T) may vary. The result should be considered negative whenever there is even a faint line.

Tenamfetamine Drug Abuse Test Kit 500ng / Ml One Test Panel Powder Diagnosis 3

POSITIVE: A colored line appears in the Control region (C) and NO line appears in the Test region (T). The positive result means that the drug concentration in the sample is greater than the designated cut-off for a specific drug.


Tenamfetamine Drug Abuse Test Kit 500ng / Ml One Test Panel Powder Diagnosis 4

 

INVALID: No line appears in the Control region (C). Insufficient specimen volume or incorrect procedural techniques are the most likely reasons for Control line failure. Read the directions again and repeat the test with a new test card. If the result is still invalid, contact your manufacturer.