Format: | AllTest | Specimen: | Oral Fluid |
---|---|---|---|
Kit Size: | 25 T/Kit | Storage: | 2-30℃ |
Cut-off: | 300 Ng/mL | Certificate: | CE |
Highlight: | rapid diagnostic test kits,home drug test kits |
Convenient And Accurate Oral Fluid Methylenedioxypyrovalerone (MDPV) Drug Abuse Diagnosis Fast Reading Test Kits With CE
Application And Description
How to use?
【DIRECTIONS FOR DEVICE USE】
Allow the test device, specimen, and/or controls to reach room temperature (15-30ºC) prior to testing. Instruct the donor to not place anything in the mouth including food, drink, gum or tobacco products for at least 10 minutes prior to collection.
1. Remove the test device from the sealed pouch and use it within one hour.
2. Take off the device cap and insert the absorbent wick to the mouth, put it under the tongue to collect oral fluid until the control line appears.
3. Put the device cap back outside of the absorbent wick and lace the test midstream on a clean and level surface. See illustration below.
4. Read results at 10 minutes. Do not read results after 15 minutes.
INTERPRETATION OF RESULTS
(Please refer to the illustration above)
NEGATIVE:* Two lines appear. One colored line should be in the control line region (C), and another apparent colored line should be in the test line region (T). A negative result indicates that the Synthetic Marijuana metabolite concentration is below the detectable level (25ng/mL).
*NOTE: The shade of color in the test line region (T) may vary, but it should be considered negative whenever there is even a faint colored line.
POSITIVE: One colored line appears in the control line region (C). No line appears in the test line region (T). A positive result indicates that the Methylenedioxypyrovalerone (MDPV) concentration exceeds the detectable level (300 ng/mL).
INVALID: Control line fails to appear. Insufficient specimen volume or incorrect procedural techniques are the most likely reasons for control line failure. Review the procedure and repeat the test using a new test. If the problem persists, discontinue using the lot immediately and contact your local distributor.
Cat. No. | Product Description | Specimen | Format | Kit Size | Cut-Off | Status |
DMDP-802 | Methylenedioxypyrovalerone (MDPV) Rapid Test Cassette | Oral Fluid | Cassette | 25 T | 300 ng/mL* | CE |
DMDP-803 | Methylenedioxypyrovalerone (MDPV) Rapid Test Device | Oral Fluid | Device | 25 T | 300 ng/mL* | CE |